Fluke1551A棒式標準溫度計 : 2399276633
Fluke1551A棒式標準溫度計
|
Thermometer probe and digital readout combined into one unit
The stainless steel probe and digital readout are fixed together and calibrated as a system. The accuracy specification
is easy to understand since it includes all uncertainty components, including drift, for up to one year1. The large backlit
LCD display rotates 90° degrees making it easy to read from any angle. A user-configurable stability/trend indicator lets
you know when your temperature is stable enough to record an accurate measurement. A user-settable auto-off function
extends typical battery life to three hundred hours. A low-battery indicator and stop-measure function prevent erroneous
measurements from being made due to low battery life. A simple three-point calibration function allows you to easily and
accuray calibrate the “Stik" Thermometer. Data logging to internal memory of up to 10,000 time-stamped measurements
is optionally available.
Why calibrate process temperature sensors?
Because temperature greatly effects the accuracy of a volumetric measurement, process manufacturers of chemical,
pharmaceutical, food or petroleum products require accurate temperature measurements—especially for processes
where quality or custody transfer are regulated by government agencies. Since all temperature sensors are subject
to drift with time, regular calibration or verification against a reliable reference thermometer is required. Finding a
reference thermometer that is accurate, repeatable and robust, however, can be challenging.
What’s wrong with my reference thermometer?
Industrial reference thermometers available on the market today, such as mercury-in-glass thermometers (or “ASTM therm
ometers") and portable electronic thermometers are useful, but both come with inherent problems. Although accurate and
repeatable, mercury thermometers are fragile. The risk of a mercury spill poses potential hazards to the environment and
to the health of personnel. Many of the United States and European Union countries have already banned their use in ind
ustrial applications. Some organizations have replaced their mercury thermometers with more durable portable electronic
thermometers, only to discover that the RTD probes lack the repeatability and dependability required of a legitimate ref
erence thermometer.
The best alternative to mercury thermometers
The Fluke “Stik" Thermometer is accurate, stable and durable, making it the best alternative to mercury thermometers and
existing electronic thermometers. The thin-film RTD sensor incorporates the same characteristics of other premium reference
thermometers designed by Hart Scientific, but it is more robust and less susceptible to drift. Accuracy and repeatability,
however, are not sacrificed for improved durability.
Using a glass-stem mercury- or spirit-filled thermometer at different immersion depths, or at significantly different ambient
temperatures from which it was calibrated, may require the application of emergent-stem corrections that can be tedious
but necessary to calculate an accurate measurement. This isn’t required with the “Stik" Thermometer. The sensor of the
1551A Ex requires a minimum immersion depth of only seven centimeters (2.8 inches), with no appreciable effect on the
measured temperature due to heat loss through stem conduction. Some digital thermometers can become less accurate
when used outside a narrow range of ambient temperatures. This isn’t the case for the “Stik" Thermometer. Your
measurement remains accurate in ambient temperatures from -10 degrees to 50 degrees celsius (14 degrees to 122 degrees
fahrenheit). With a superior probe design combined with electronic functions that assist accurate measurement, the “Stik"
Thermometer surpasses the capabilities of other digital thermometers and is the perfect substitute for your mercury
thermometers. 1In order to minimize sensor drift, the probe should always be protected from mechanical shock.
Five reasons to replace mercury thermometers
Specifications | 1551A Ex | 1552A Ex |
Temperature range | -50 °C to 160 °C | -80 °C to 300 °C |
Accuracy (1 year) | ± 0.05 °C (± 0.09 °F) | |
Display units | °C, °F | |
Sensor type | 100 Ω thin-film RTD | 100 Ω wire-wound PRT |
Probe temperature coefficient | 0.00385 Ω/Ω/°C nominal | |
Sensor length | ≤ 10 mm | ≤ 30 mm |
Sensor position (from sheath tip) | 3 mm (0.1 in) | |
Minimum immersion depth | 7 cm (2.8 in) | 12 cm (4.8 in) |
Probe sheath material | Stainless steel | |
Response time | 4.8 mm diameter probe (3/16 in): 14 seconds | |
Probe hysteresis | ± 0.01 °C | |
Temperature resolution | Selectable: 0.1, 0.01, 0.001 (default 0.01) | |
Sample rate | Selectable: 0.5 seconds, 1 seconds, 2 seconds (default 1 seconds) | |
Operating temperature range readout | -10 °C to 50 °C (14 °F to 122 °F) | |
Temperature resolution | Selectable: 0.1, 0.01, 0.001 (default 0.01) | |
Sample rate | Selectable: 0.5 seconds, 1 seconds, 2 seconds (default 1 seconds) | |
Operating temperature range readout | -10 °C to 50 °C (14 °F to 122 °F) | |
Storage temperature | -20 °C to 60 °C (-4 °F to 140 °F) 0 % to 95 % RH, non-condensing | |
Optional data logging1 | Up to 10,000 time-stamped readings stored to internal memory | |
Logging intervals1 | 2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds; | |
Damping | Moving average of the most recent 2 to 10 readings | |
Communications | RS-232 stereo jack (access calibration parameters only) | |
AC power | None | |
DC power | 3-AAA Batteries, typical battery life of 300 hours without LCD backlight | |
EMC compliance | EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class “B | |
Enclosure rating | IP50 | |
Electronics dimensions (H x W x D) | 114 mm x 57 mm x 25 mm | |
Weight | 196 g (6.9 oz) | |
Calibration (included) | NVLAP-accredited, NIST-traceable | |
Characterization | CVD | ITS-90 |
型號名稱 | 描述 |
1551A-9 | 1551A Ex Thermometer, Fixed RTD, -50 °C to 160 °C, 4.8 mm x 229 mm (3/16 in x 9 in) Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
1551A-12 | 1551A Ex Thermometer, Fixed RTD, -50 °C to 160 °C, 6.35 mm x 305 mm (1/4 in x 12 in) Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
1551A-20 | 1551A Ex Thermometer, Fixed RTD, -50 °C to 160 °C, 6.35 mm x 508 mm (1/4 in x 20 in) Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
1551A-9-DL | 1551A Ex Datalogging options Thermometer, Fixed RTD, -50 °C to 160 °C, 4.8 mm x 229 mm (3/16 in x 9 in) with Datalog Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
1551A-12-DL | 1551A Ex Datalogging options Thermometer, Fixed RTD, -50 °C to 160 °C, 6.35 mm x 305 mm (1/4 in x 12 in) with Datalog Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
1551A-20-DL | 1551A Ex Datalogging options Thermometer, Fixed RTD, -50 °C to 160 °C, 6.35 mm x 508 mm (1/4 in x 20 in) with Datalog Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
1552A-12 | Thermometer, Fixed PRT, -80 °C to 300 °C, 6.35 mm x 305 mm (1/4 in x 12 in) Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
1552A-12-DL | Thermometer, Fixed PRT, -80 °C to 300 °C, 6.35 mm x 305 mm (1/4 in x 12 in) with Datalog Includes NVLAP-accredited report of calibration, User’s Guide on CD-ROM, 3 AAA batteries |
Note: Model number appended with a -9, -12, -20 indicates probe sheath length in inches. All probe diameters are 6.35 mm (1/4 in.) with the exception of the 1551A-9, which is 4.8 mm diameter (3/16 in.). |
機套和掛件 | |
1551-CASE | 1551-CASE |
電纜和硬件 | |
1551-CBL | Cable, RS-232 155X |
您感興趣的產品PRODUCTS YOU ARE INTERESTED IN