污水處理設備 污泥處理設備 水處理過濾器 軟化水設備/除鹽設備 純凈水設備 消毒設備|加藥設備 供水/儲水/集水/排水/輔助 水處理膜 過濾器濾芯 水處理濾料 水處理劑 水處理填料 其它水處理設備
廈門慧嘉生物科技有限公司
暫無信息 |
閱讀:422發布時間:2014-4-16
大鼠糖化血紅蛋白A1c(GHbA1c)ELISA試劑盒使用說明
1
Rat glycated hemoglobin A1c (GHbA1c) ELISA Kit
Catalog Number. CSB-E08140r
For the quantitative determination of rat glycated hemoglobin A1c
(GHbA1c) concentrations in lysate for RBC.
This package insert must be read in its entirety before using this product.
In order to obtain higher efficiency service, please ready to supply the lot number
of the kit to us (found on the outside of the box).2
PRINCIPLE OF THE ASSAY
This assay employs the direct competitive inhibition enzyme immunoassay
technique. Antibody specific for GHbA1c has been pre-coated onto a microplate.
Standards and samples are pipetted into the wells with biotin-conjugated
GHbA1c. A competitive inhibition reaction is launched between GHbA1c
(Standards or samples) and biotin-conjugated GHbA1c with the pre-coated
GHbA1c antibody. After washing, avidin conjugated Horseradish Peroxidase
(HRP) is added to the wells. Following a wash to remove any unbound reagent, a
substrate solution is added to the wells and color develops in opposite to the
amount of GHbA1c bound in the initial step. The color development is stopped
and the intensity of the color is measured.
DETECTION RANGE
1.25 ng/ml-80 ng/ml.
SENSITIVITY
The minimum detectable dose of rat GHbA1c is typically less than 0.625 ng/ml.
The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as
the lowest rat GHbA1c concentration that could be differentiated from zero. It
was determined the mean O.D value of 20 replicates of the zero standard added
by their three standard deviations.
SPECIFICITY
This assay has high sensitivity and excellent specificity for detection of rat
GHbA1c. No significant cross-reactivity or interference between rat GHbA1c and
analogues was observed.
Note: Limited by current skills and knowledge, it is impossible for us to complete
the cross-reactivity detection between rat GHbA1c and all the analogues,
therefore, cross reaction may still exist.3
PRECISION
Intra-assay Precision (Precision within an assay): CV%<8%
Three samples of known concentration were tested twenty times on one plate to
assess.
Inter-assay Precision (Precision between assays): CV%<10%
Three samples of known concentration were tested in twenty assays to assess.
LIMITATIONS OF THE PROCEDURE
? FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC
PROCEDURES.
? The kit should not be used beyond the expiration date on the kit label.
? Do not mix or substitute reagents with those from other lots or sources.
? If samples generate values higher than the highest standard, dilute the
samples with Sample Diluent and repeat the assay.
? Any variation in Sample Diluent, operator, pipetting technique, washing
technique, incubation time or temperature, and kit age can cause variation
in binding.
? This assay is designed to eliminate interference by soluble receptors,
binding proteins, and other factors present in biological samples. Until all
factors have been tested in the Immunoassay, the possibility of
interference cannot be excluded.4
MATERIALS PROVIDED
Reagents Quantity
Assay plate 1(96 wells)
Standard (100 x concentrate) 1 x 100 μl
Biotin-conjugate (100 x concentrate) 1 x 60 μl
HRP-avidin (100 x concentrate) 1 x 120 μl
Biotin-conjugate Diluent 1 x 10 ml
HRP-avidin Diluent 1 x 20 ml
Sample Diluent 2 x 20 ml
Wash Buffer (25 x concentrate) 1 x 20 ml
TMB Substrate 1 x 10 ml
Stop Solution 1 x 10 ml
Adhesive Strip (For 96 wells) 4
Instruction manual 1
STORAGE
Unopened
kit
Store at 2 - 8°C. Do not use the kit beyond the expiration date.
Opened kit
Coated assay
plate
May be stored for up to 1 month at 2 - 8°C.
Try to keep it in a sealed aluminum foil bag,
and avoid the damp.
Standard May be stored for up to 1 month at 2 - 8° C.
If don’t make recent use, better keep it store
at -20°C.
HRP-avidin
Biotin-conjugate
Biotin-conjugate
Diluent
May be stored for up to 1 month at 2 - 8°C.
HRP-avidin
Diluent
Sample Diluent
Wash Buffer
TMB Substrate
Stop Solution
*Provided this is within the expiration date of the kit.5
OTHER SUPPLIES REQUIRED
? Microplate reader capable of measuring absorbance at 450 nm, with the
correction wavelength set at 540 nm or 570 nm.
? An incubator which can provide stable incubation conditions up to
37°C±0.5°C.
? Squirt bottle, manifold dispenser, or automated microplate washer.
? Absorbent paper for blotting the microtiter plate.
? 100ml and 500ml graduated cylinders.
? Deionized or distilled water.
? Pipettes and pipette tips.
? Test tubes for dilution.
PRECAUTIONS
The Stop Solution provided with this kit is an acid solution. Wear eye, hand, face,
and clothing protection when using this material.6
SAMPLE COLLECTION AND STORAGE
? Lysate for RBC Collect a certain volume of anticoagulant blood sample
(e.g. 1ml). Centrifuge at 10000rpm for 10min at 4°C to collect the RBC.
Remove supernate and add cold pure water to redissolve RBC to the
previous volume (e.g. 1ml). Then store at -20°C for about 30-60 minutes.
After three freeze-thaw cycles to break up the cell membranes, the cell
lysates were centrifuged for 10 minutes at 10000rpm, 4°C. Collect the
supernatant. RBC lysates should be assayed immediay or aliquotted
and stored at -20°C. Centrifuge the sample again after thawing before the
assay. Avoid repeated freeze-thaw cycles.
SAMPLE PREPARATION
? Recommend to dilute the lysate for RBC with Sample Diluent(1:5000)
before test. The suggested 5000-fold dilution can be achieved by adding
2μl sample to 98μl of Sample Diluent. Complete the 5000-fold dilution by
adding 3μl of this solution to 297μl of Sample Diluent. The recommended
dilution factor is for reference only. The optimal dilution factor should be
determined by users according to their particular experiments.7
Note:
1. CUSABIO is only responsible for the kit itself, but not for the samples
consumed during the assay. The user should calculate the possible
amount of the samples used in the whole test. Please reserve sufficient
samples in advance.
2. If the samples are not indicated in the manual, a preliminary experiment to
determine the validity of the kit is necessary.
3. Please predict the concentration before assaying. If values for these are
not within the range of the standard curve, users must determine the
optimal sample dilutions for their particular experiments.
4. Owing to the possibility of mismatching between antigen from other
resource and antibody used in our kits (e.g., antibody targets
conformational epitope rather than linear epitope), some native or
recombinant proteins from other manufacturers may not be recognized by
our products.
5. Influenced by the factors including cell viability, cell number and also
sampling time, samples from cell culture supernatant may not be detected
by the kit.
6. Fresh samples without long time storage are recommended for the test.
Otherwise, protein degradation and denaturalization may occur in those
samples and finally lead to wrong results.8
REAGENT PREPARATION
Note:
? Kindly use graduated containers to prepare the reagent. Please don't
prepare the reagent directly in the Diluent vials provided in the kit.
? Bring all reagents to room temperature (18-25°C) before use for 30min.
? Prepare fresh standard for each assay. Use within 4 hours and discard
after use.
? Making serial dilution in the wells directly is not permitted.
? Distilled water is recommended to be used to make the preparation for
reagents or samples. Contaminated water or container for reagent
preparation will influence the detection result.
1. Biotin-conjugate (1x) - Centrifuge the vial before opening.
Biotin-conjugate requires a 100-fold dilution. A suggested 100-fold
dilution is 10 μl of Biotin-conjugate + 990 μl of Biotin-conjugate Diluent.
2. HRP-avidin (1x) - Centrifuge the vial before opening.
HRP-avidin requires a 100-fold dilution. A suggested 100-fold dilution is 10
μl of HRP-avidin + 990 μl of HRP-avidin Diluent.
3. Wash Buffer(1x)- If crystals have formed in the concentrate, warm up to
room temperature and mix gently until the crystals have compley
dissolved. Dilute 20 ml of Wash Buffer Concentrate (25 x) into deionized or
distilled water to prepare 500 ml of Wash Buffer (1 x).9
4. Standard
Centrifuge the standard vial at 6000-10000rpm for 30s.
Dilute the Standard(100x) with Sample Diluent. A suggested 100-fold
dilution is 5 μl of Standard(100x) + 495 μl of Sample Diluent. This diluted
Standard (S7) serves as the high standard (80 ng/ml).
Do not substitute other diluents. Mix the standard to ensure complete
dilution and allow the standard to sit for a minimum of 15 minutes with
gentle agitation prior to making dilutions.
Pipette 150 μl of Sample Diluent into each tube (S0-S6). Use the stock
solution to produce a 2-fold dilution series (below). Mix each tube
thoroughly before the next transfer. Sample Diluent serves as the zero
standard (0 ng/ml).
Tube S7 S6 S5 S4 S3 S2 S1 S0
ng/ml 80 40 20 10 5 2.5 1.25 010
ASSAY PROCEDURE
Bring all reagents and samples to room temperature before use. Centrifuge
the sample again after thawing before the assay. It is recommended that all
samples and standards be assayed in duplicate.
1. Prepare all reagents, working standards, and samples as directed in the
previous sections.
2. Refer to the Assay Layout Sheet to determine the number of wells to be
used and put any remaining wells and the desiccant back into the pouch
and seal the ziploc, store unused wells at 4°C.
3. Set a Blank well without any solution.
4. Add 50μl of standard and sample per well.
5. Add 50μl of Biotin-conjugate (1x) to each well(not to Blank well). Cover
with a new adhesive strip. Incubate for 40 minutes at 37°C.
(Biotin-conjugate (1x) may appear cloudy. Warm up to room temperature
and mix gently until solution appears uniform.)
6. Aspirate each well and wash, repeating the process two times for a total of
three washes. Wash by filling each well with Wash Buffer (200μl) using a
squirt bottle, multi-channel pipette, manifold dispenser, or autowasher,
and let it stand for 2 minutes, complete removal of liquid at each step is
essential to good performance. After the last wash, remove any remaining
Wash Buffer by aspirating ordecanting. Invert the plate and blot it against
clean paper towels.
7. Add 100μl of HRP-avidin (1x) to each well(not to Blank well). Cover the
microtiter plate with a new adhesive strip. Incubate for 40 minutes at 37°C.
8. Repeat the aspiration/wash process for five times as in step 6.
9. Add 90μl of TMB Substrate to each well. Incubate for 20 minutes at 37°C.
Protect from light.
10. Add 50μl of Stop Solution to each well, gently tap the plate to ensure
thorough mixing.11
11. Determine the optical density of each well within 5 minutes, using a
microplate reader set to 450 nm. If wavelength correction is available, set
to 540 nm or 570 nm. Subtract readings at 540 nm or 570 nm from the
readings at 450 nm. This subtraction will correct for optical imperfections
in the plate. Readings made directly at 450 nm without correction may be
higher and less accurate.
*Samples may require dilution. Please refer to Sample Preparation section.
Note:
1. The final experimental results will be closely related to validity of the
products, operation skills of the end users and the experimental
environments.
2. Samples or reagents addition: Please use the freshly prepared Standard.
Please carefully add samples to wells and mix gently to avoid foaming. Do
not touch the well wall as possible. For each step in the procedure, total
dispensing time for addition of reagents or samples to the assay plate
should not exceed 10 minutes. This will ensure equal elapsed time for
each pipetting step, without interruption. Duplication of all standards and
specimens, although not required, is recommended. To avoid
cross-contamination, change pipette tips between additions of each
standard level, between sample additions, and between reagent additions.
Also, use separate reservoirs for each reagent.
3. Incubation: To ensure accurate results, proper adhesion of plate sealers
during incubation steps is necessary. Do not allow wells to sit uncovered
for extended periods between incubation steps. Once reagents have been
added to the well strips, DO NOT let the strips DRY at any time during the
assay. Incubation time and temperature must be observed.
4. Washing: The wash procedure is critical. Complete removal of liquid at
each step is essential to good performance. After the last wash, remove
any remaining Wash Solution by aspirating or decanting and remove any
drop of water and fingerprint on the bottom of the plate. Insufficient
washing will result in poor precision and falsely elevated absorbance
reading. When using an automated plate washer, adding a 2 minutes soak
period following the addition of wash buffer, and/or rotating the plate 180
degrees between wash steps may improve assay precision.12
5. Controlling of reaction time: Observe the change of color after adding TMB
Substrate (e.g. observation once every 10 minutes), TMB Substrate
should change from colorless or light blue to gradations of blue. If the color
is too deep, add Stop Solution in advance to avoid excessively strong
reaction which will result in inaccurate absorbance reading.
6. TMB Substrate is easily contaminated. TMB Substrate should remain
colorless or light blue until added to the plate. Please protect it from light.
7. Stop Solution should be added to the plate in the same order as the TMB
Substrate. The color developed in the wells will turn from blue to yellow
upon addition of the Stop Solution. Wells that are green in color indicate
that the Stop Solution has not mixed thoroughly with the TMB Substrate.13
ASSAY PROCEDURE SUMMARY
*Samples may require dilution. Please refer to Sample Preparation section.14
CALCULATION OF RESULTS
Using the professional soft "Curve Expert 1.3" to make a standard curve is
recommended, which can be downloaded from our web.
Average the duplicate readings for each standard and sample and subtract the
average optical density of Blank.
Create a standard curve by reducing the data using computer software capable
of generating a four parameter logistic (4-PL) curve-fit. As an alternative,
construct a standard curve by plotting the mean absorbance for each standard
on the x-axis against the concentration on the y-axis and draw a best fit curve
through the points on the graph. The data may be linearized by plotting the log of
the GHbA1c concentrations versus the log of the O.D. and the best fit line can be
determined by regression analysis. This procedure will produce an adequate but
less precise fit of the data.
If samples have been diluted, the concentration read from the standard curve
must be multiplied by the dilution factor.
慧嘉生物您實驗身邊的好伙伴
為客戶提供“zui高品質的產品”和“zui的服務”
詳細說明書及技術咨詢
:
陳
:1193953234 1048735792
環保在線 設計制作,未經允許翻錄必究 .? ? ?
請輸入賬號
請輸入密碼
請輸驗證碼
請輸入你感興趣的產品
請簡單描述您的需求
請選擇省份